± 0.1 % of Span or ± 0.3° C (± 0.54° F) whichever is greater (Including Linearization Error)
Relative Error within ± 0.2 % between Output-1 & 2 - Dual Output
Burn-Out
60 seconds or Less - Burn-Out Time
Down
Off
Up
Consumption
100 mA at 24 Vdc (WRUA) - Current Consumption
11 VA at 100 Vac (WRUA) - Power Consumption
7 VA at 100 Vac (WRUV) - Power Consumption
70 mA at 24 Vdc (WRUV) - Current Consumption
Electrical Connectors
M4 Screw Terminal
Enclosure, Body Material
ABS Resin
Humidity
5 to 90 % RH, Non-Condensing - Operating Humidity Range
Insulation Resistance
100 MΩ or more at 500 Vdc between Input & Output, Input & Power Supply, Input & Ground, Output & Power Supply, Output & Ground, & Power Supply & Ground
Mounting
DIN Rail Mounting
Rack-Mounted
Wall Mounted
Operating Temperature
0° to 50° C (32° to 122° F)
Output
0-24 mA
10 mV - 20 V - Span (DC)
-10 to 10 Vdc
1-24 mA - Span (DC)
Power Supply
24 Vdc ± 10 % (DC Drive)
85-264 Vac, 47 to 63 Hz (AC Drive)
Power Supply Effect
± 0.1 % of Span or 0.3° C (0.54° F) or Less - Power Supply Voltage Fluctuations
Resistance
1 MΩ - Input Resistance During Power-On
2 to 250 kΩ or more - Allowable Load Resistance for DC Voltage Output
3 kΩ - Input Resistance During Power Off
50 ohms or Less Per Wire - Allowable Leadwire Resistance
750 to 15,000 ohms or Less - Allowable Load Resistance for DC Current Output
Response Time
200 ms, 63 % Response (10 to 90 %)
Span & Zero Adjustments
± 1 % - Input Adjustment
± 10 % - Output Adjustment
Supply Voltage
1,500 Vac/minute for AC Drive - Withstand Voltage
500 Vac/minute between Output & Power Supply - DC Drive Withstand Voltage
Temperature Effect
± 0.2 % of Span or 0.6° C (1.08° F) or Less for a Change of 10° C (18° F) - Ambient Temperature Change