Accuracy
- ± 0.025 % of Full Scale
- ± 0.03 % of Full Scale
- ± 0.10 % of Full Scale
- AM2-1 Low Pressure Module & AM2-2 Medium / High Pressure Module: 0.1 % of Span (10,000 psi Only Offered in psig & 0.1 % Accuracy)
- AM2-2 Medium / High Pressure Module: 0.05 % of Span
Ambient Temperature Range
- AM2-2 Medium / High Pressure Module: -6.6° to 48.8° C (20° to 120° F)
Approvals & Certifications
Battery Life
- Up to 40 Hours Battery with 2 Modules Installed
Compensated Temperature
- AM2-1 Low, Medium / High Pressure Module: -7° to 49° C (20° to 120° F)
Media
- AM2-1 Low Pressure Module: Clean, Non-Corrosive Dry Gases
Operating Temperature
- -20° to 70° C (-4° to 158° F)
Overpressure Effect
- Overpressure Effect: 150 % for Ranges to 1,000 psi & 200 % for Ranges > 1,000 psi
Power Requirement
- AA Batteries
- USB Universal AC Adapter: 100-240 Vac, 50/60 Hz
Power Supply
- Loop: 24 Vdc at 35 mA - Max.
Pressure Type
- Compound Pressure
- Differential
- Gauge
Repeatability
- AM2-1 Low Pressure Module: ± 0.02 % of Span (Range Below 1" H₂O)
- AM2-1 Low, Medium / High Pressure Module: ± 0.01 % of Span (Range 0 / 1" H₂O or Higher)
Resolution
- ± 0.001 % of Span
- 0.001 mV
- 99,999 Counts
- AM2-RT1 & RT2 Module: 0.004 ohm
- AM2-RT1 & RT2 Module: 0.01° C or F
- AM2-RT1 Module: 0.1° C
Sensitivity
- AM2-1 Low, Medium / High Pressure Module: ± 0.002 % of Span Typical
- AM2-2 Medium / High Pressure Module: ± 0.001 % of Span, Max.
Static Pressure
- AM2-1 Low Pressure Module, Max.
Storage Temperature
- -20° to 70° C (-4° to 158° F)
Temperature Effect
- ± 0.001 % of Span/°F over Compensated Range from Reference Temperature of 70° F (± 3°)
- AM2-1 Low, Medium / High Pressure Module: ± 0.004 % of Span/°F over Compensated Range from Reference Temperature Range of 70° F (± 3°)
Update Time
- Warm-Up Time: 5 minutes for Rated Accuracy