Specifications
Current
- 40-60 mA - Current Control Circuit
Dissipation
- 100 mA, 24 Vdc - Current Dissipation
- 20 mA, 110 Vdc - Current Dissipation
- 6 VA, 100 Vdc - Power Dissipation
- 9 VA, 200 Vac - Power Dissipation
Duty Cycle
- Within 50 ± 30 % when 100 % Input - Duty Cycle for Input Pulse Width
Frequency
- 0 to 100 Hz (Pulse Width more than 3 ms)
- 0.1 Hz to 10 kHz - Input Frequency
Humidity
- 5 to 90 % RH, Non-Condensing
Input Signal
- -1 to 8 V - Low Level Voltage Pulse
- 2-24 V - High Level Voltage Pulse
Insulation Resistance
- Over 100 MΩ (500 Vdc) between Input, Output, Power Supply, Ground
Mounting
- DIN Rail Mounting
- Wall Mounted
Output Signal
- 0-10 Vdc
- 0-5 mA DC
- 0-100 mVdc
- 0-20 mA DC
Power Supply
- 12 Vdc ± 10 % - Transmitter Power Supply
- 170-264 Vac (47 to 63 Hz)
- 24 Vdc ± 10 % - Transmitter Power Supply
- 85-132 Vac (47 to 63 Hz)
- 85-150 Vdc
Power Supply Effect
- Less than ± 0.1 % of Span for Fluctuation of 24 Vdc ± 10 %, 85-132 Vac (47 to 63 Hz) / 85-150 Vdc or 170-264 Vac - Effect of Power Voltage Fluctuation
Resistance
- 1 Kiloohm (When Current Pulse Input) - Input Resistance
- 1 to 100 ohms or Less - Output Resistance
- 10 kΩ or more - High Level: Dry Voltage Contact
- 2 to 250 kΩ or more - Permissible Load Resistance
- 200 ohms (When Current Pulse Input) - Input Resistance
- 200 ohms or Less - Low Level: Dry Voltage Contact
- 500 kΩ or more - Output Resistance
- 500 ohms (When Current Pulse Input) - Input Resistance
- More Than 10 kΩ (When Voltage Pulse Input) - Input Resistance
Span & Zero Adjustments
- ± 10 % of Span - Output Adjust Range
Supply Voltage
- 1,500 Vac/minute between Input, Output, Power Supply, Ground - Withstand Voltage
Temperature Effect
- Less than ± 0.2 % of Span for 10° C (50° F) Temperature Range - Effect of Ambient Temperature Change
Temperature Range
- 0° to 50° C (32° to 122° F)