Specifications
Accuracy
- ± 0.2 mg/L or ± 2 % saturation within ± 10° C of the calibration Temperature within the Range 0° to 35° C
Range
- Dissolved oxygen: 0 to 20 ppm
Material of Construction
- Glass-Coupled Polypropylene
Response Time
- Typically 20 seconds for 90 % step change of oxygen concentration at 20° C
Process Connection
- 1/2" BSP
- 1" BSP
- 1/2" NPT
- 1" NPT
Cable Length
- Standard 1 to 20 m
- Maximum 100 m with junction box